Technical Specifications –
Flame System –
Instrument can be configured with Flame, Graphite or Integrated Flame & Graphite (interchanged using Software)
| Parameter | Value |
|---|---|
| Wavelength Range | 185nm – 915nm |
| Optics | Double Beam, Sealed & Vibration free optical System with a reflective optical compartment |
| Detector | PMT (Wide Range) |
| Monochromator | Czerny-turner type with Holographic Grating of 1800 lines/mm and Reciprocal Linear Dispersion better than 1.6 nm/mm, focal length 300mm. |
| Spectral Bandwidth | Adjustable slit from 0.1 nm. to 2 nm. (software selectable) |
| Wavelength Accuracy | ± 0.15nm |
| Wavelength Reproducibility | < 0.05nm |
| Resolution | 0.2nm ± 0.02nm |
| Baseline Stability | 0.005A/30min |
| Sensitivity (Cu) | >0.9 Absorbance or better for 5 ppm Cu solution-12 |
| Detection Limit | Cu < 0.004 μg/ml (flame) Cd < 0.4 x 10 g (graphite furnace) |
| Repeatability | Cu < 0.7% ( Air/Acetylene flame) Ba < 1.0% ( Nitrous oxide/Acetylene flame) Cu < 2.0% Cd < 2.0% (Graphite Furnace) |
| Background Correction | Deuterium Arc / Self reversal upto 2.5 Absorbance |
| Characteristic Concentration | Cu < 0.02 μg/ml, Ba < 0.15 mg/ml (NO/Acetylene) |
| Burner Heads | 2 |
| Nebuliser | High Efficiency Nebuliser |
| Burner Head | Titanium Alloy |
| Atomization Chamber | Corrosion-resistant material |
| Position Adjustment | Automatic changeover (AA8000GF) Manual (AA8000F) Automatic Setting of Optimum Height for Flame Burner. Burner movement in horizontal & vertical direction by software. |
| Safety Functions (Interlocks) | Burner Identification, Flame Sensor, Gas leak Sensor, Low Gas Pressure Sensor, Drain Trap Sensor, Power Loss Protection, Circulation Water (graphite), Over Temperature Sensor (graphite) |
Graphite System
- Graphite Tubes are pyrolytically coated & transversely heated upto 3000˚C
- Platform tubes supplied as standard, capable of accepting volumes upto 20 µL
- Programmable graphite furnace upto 10 stages which can be set & stored using Software
- Replacement of Graphite tube performed by a simple command using Software
- Graphite tube cooling by additional water circulatory system
- Cd Sensitivity: 4×10 gm
Safety Features –
- Argon Gas pressure sensor
- Water Flow sensor
- Over Temperature sensor
- Broken Graphite Tube protection